Fluke Calibration 4180 Precision IR Calibrators
The Fluke 4180 Precision Infrared Calibrator is designed for accuracy and traceability in infrared temperature measurements. Now it’s easy to increase your IR temperature measurement accuracy with the new fluke 4180 Hart Scientific Precision Infrared Calibrators from Fluke Calibration.
These calibrators give more consistent, accurate and reliable calibrations because emissivity is radiometrically calibrated, and the target size minimizes size of source effect errors. In addition, these calibrators simplify calibration because they uniquely compensate for errors caused by thermometer emissivity settings.
New accredited performance for point-and-shoot calibrations
- Calibrated radiometrically for meaningful, consistent results.
- Accredited calibration included
- Accurate, reliable performance from –15 °C to 500 °C
- Large target size of 152 mm (6 in) is required for calibrating most thermometers.
- Light instrument weight of 8.6 kg (19 lbs) makes it easy to lift and carry.
- Intuitive, easy-to-read display that indicates when temperature is stable.
Fluke Calibration 4180 Precision IR Calibrators Specifications:
Fluke Calibration 4180:
- Temperature range (@ 23°C ambient, 0.95 emissivity):–15°C to 120°C
- Display accuracy:±0.40°C at –15°C,±0.40°C at 0°C,±0.50°C at 50°C,±0.50°C at 100°C,±0.55°C at 120°C
- Stability:±0.10°C at –15°C,±0.05°C at 0°C,±0.10°C at 120°C
- Uniformity (5.0 in dia of center of target):±0.15°C at –15°C,±0.10°C at 0°C,±0.25°C at 120°C
- Uniformity (2.0 in dia of center of target):±0.10°C at -15°C,±0.10°C at 0°C,±0.20°C at 120°C
- Heating time:15 min: –15°C to 120°C,14 min: 23°C to 120°C
- Cooling time:15 min: 120°C to 23°C,20 min: 23°C to –15°C
- Target diameter: 152.4 mm (6 in)
- Power:115 V AC (±10%), 6.3 A,50/60 Hz, 630 W,230 V AC (±10%), 3.15 A,,50/60 Hz, 630 W
Fluke Calibration 4181:
- Temperature range (@ 23°C ambient, 0.95 emissivity):
- Display accuracy:±0.35°C at 35°C,±0.50°C at 100°C,±0.70°C at 200°C,±1.20°C at 350°C,±1.60°C at 500°C
- Stability:±0.05°C at 35°C,±0.20°C at 200°C,±0.40°C at 500°C
- Uniformity (5.0 in dia of center of target):±0.10°C at 35°C,±0.50°C at 250°C,±1.00°C at 500°C
- Uniformity (2.0 in dia of center of target):±0.10°C at 35°C,±0.25°C at 250°C,±0.50°C at 500°C
- Heating time:45 min: 35⁰ C to 500⁰ C
- Cooling time:100 min: 500°C to 35°C,40 min: 500°C to 100°C
- Target diameter:152.4 mm (6 in)
- Power:115 V AC (±10%), 10 A,50/60 Hz, 1000 W,230 V AC (±10%), 5 A,,50/60 Hz, 1000 W