Kyoritsu 5050-00 Ior Logger
The Kyoritsu 5050-00 Ior Logger is a powerful tool for measuring and recording leakage current, giving vital data for safety and maintenance. This advanced leakage current logger is designed to detect electrical leaks in diverse wiring systems. It also measures and logs Ior up to 4 parameters with a measured value that is less susceptible to harmonics.
The ability of the Kyoritsu 5050-00 Ior Logger to correctly measure and record leakage current is one of its key advantages. This is critical in a variety of industries, including electrical, medical, and automotive, where high leakage current can be hazardous or even fatal. The device is also incredibly simple to use, with a basic interface that allows users to monitor data and modify settings quickly and effortlessly. Furthermore, the Kyoritsu 5050-00 Ior Logger is small and portable, making it simple to transport from one location to another.
Kyoritsu 5050-00 Ior Logger, 4 channels Features
- Measures and logs up to four channels at the same time.
- Different wiring schemes (single-phase 2 & 3-wire, three-phase 3 & 4-wire) are supported.
- The world’s fastest 200 ms interval for measuring leakage current
- Provides typical leakage and load current measurements.
- A large graphic display and a magnet on the back casing allow it to be attached to metal enclosures.
Kyoritsu 5050-00 Ior Logger specifications:
- Measurements and Parameters Ior: Leakage current (Trms) with resistive components only
Io: Leakage current (Trms) with a basic wave of 40 to 70 Hz
Iom: Leakage current (Trms) including harmonic components
V: Reference voltage (Trms) with a basic wave of 40 to 70 Hz
Vm: Reference voltage (Trms) including harmonic components
R : Insulation resistance, Frequency(Hz), Phase angle
- Other Functions: Digital output, print screen, backlight, data hold
- Recording Interval: 200/400 ms/1/5/15/30 s/1/5/15/30 m/1/2 hours
- Ior Range: 10/100/1000 mA/10 A/AUTO
- Voltage Range: 1000 V
- Phase Angle Display Range: 0 to ±180° (regarding the phase of reference voltage as 0°)
- Frequency Meter Range: 40 to 70 Hz
- Power Source: 6 x 1.5 AA LR6 (Battery life approx. 11h)
- Applicable Standards: EC61010-1 CATIV 300V , CATIII 600V Pollution degree 2
IEC61010-2-030 , IEC61010-031 , IEC61326